Loading...
Andersen Products offer sterilisation services for health professionals, medical device manufacturers and veterinary practices. Our labs are located in Clacton on Sea in Essex and we can offer our services to customers throughout Europe.
Your devices are sterilised according to your parameters in one of our EOGas systems.
30 °C cycle – 12 or 24 hours
50 °C cycle – 3 or 5 hours |
1 x 40 l bag |
Aeration at 30/50 °C in the steriliser or at room temperature in dedicated aeration room. |
50 °C cycle – 16 hours
|
Up to 10 x 60 l bag |
Aeration at room temperature in dedicated aeration room.
|
In all cases, aeration is continued for 48 hours, whilst our biological indicators (BIs) are incubating.
SAL 10-6 delivered
Your goods are returned with a certificate of conformity.
You can sterilise almost anything, except liquid, food and drugs! Please visit our section Items suitable for sterilisation with EtO to find out what you can send to us for sterilisation.